Back to Article

Magnetic-flux-density based rational designing of 100kHz electromagnetic-field radiator for nano-thermal ablation therapy

World Journal of Cancer and Oncology Research | Vol 5, Issue 1

Table 1. Target cancers, radiation conditions and medical findings

Pts # Cancer Irradiation site Anesthesia Output (kW) Clearance to skin (mm) Density at skin (mT) Skin temperature initial/end Medical findings
Interview Diagnosis
1 melanoma III 65 male Right groin one local const 0.9 0 19.9 34.8 / 35.0 No ache and heat feeling during and after treatment No burn
2 melanoma (meta) IV 65 female left side breast A local avg 4.8 1 mm 27.3 36.1 / 37.1 No ache and heat feeling during and after treatment No burnNo abnormality of blood pressure and heart rate
B 27.3 NT
3 melanoma (meta) IIIc 38 female left inner thigh A none const 7.4 1 mm 34.9 36.0 / 37.3 No ache and heat feeling during and after treatmentSlight tingling sensation during irradiation No burnNo abnormality of blood pressure and heart rate
B 34.9 NT
C 34.9 NT
4 papillary throid IV (meta) 80 female neck under chin one local const 6.8 13 mm 19 34.0 / 35.5 No ache and heat feeling during and after treatment No burn, No abnormality Permittance of bathing
5 brest (meta) 80 female right side chest one none avg 3.5 5 mm 20.4 35.3 / 37.0 same as above same as above
6 papillary throid (meta) 45 female right front arm one local const 7.2 3 mm 32.1 NT same as above same as above
7 tongue cancer (meta) chest under chin one local const 8.4 10 mm 23.2 NT same as above same as above