Quantifying 64 drugs, illicit substances, and D- and L- isomers in human oral fluid with liquid-liquid extraction

Brian Robbins1, Rob E. Carpenter2,*, Mary Long1,3 and Jacob Perry4
1
Department of Research, Advanta Genetics, USA
2
Department of Human Resource Development, University of Texas at Tyler, USA
3
University of Alabama at Birmingham, USA
4
University of Miami Miller School of Medicine, USA

Abstract

Although human oral fluid has become more routine for quantitative drug detection in pain management, detecting a large scope of medications and substances is costly and technically challenging for laboratories. This paper presents a quantitative assay for 64 pain medications, illicit substances, and drug metabolites in human oral fluid. The novelty of this assay is that it was developed on an older model AB SCIEX 4000 instrument and renders obscure the need for more technical and expensive laboratory equipment. This method includes addition of internal standard and a 2-step liquid-liquid extraction and dry-down step to concentrate and clean the samples. The samples were suspended in 50% MeOH in water and separation and detection was accomplished using triple quadrupole mass spectrometry (LC-MS/MS). Separation was achieved using reverse-phase liquid chromatography with detection by LC-MS/MS. A second injection was done in negative mode to determine THC-COOH concentration as an indicator of THC. An aliquot of the (already) extracted samples was analyzed for D- and L- isomers of amphetamine and methamphetamine using a chiral column. The standard curve spanned from 5 to 2000 ng/mL for most of the analytes (1 to 2000 ng/mL for fentanyl and THC-COOH) and up to 1000 ng/mL for 13 analytes. Pregabalin and gabapentin ranged from 25 to 2000 ng/mL. The result is a low-cost method for the sensitive detection of a wide-ranging oral fluid menu for pain management. This assay has a high sensitivity, and good precision and accuracy for all analytes with an older model mass spectrometer.

1. Introduction

The United States Controlled Substances Act (CSA) places drugs and certain chemicals used to make drugs into five distinct categories or schedules depending upon the drug’s acceptable medical use and its potential for abuse or dependency [1]. The abuse rate is a determinate factor in the scheduling of the drug; for example, Schedule I drugs are considered the most dangerous class of drugs with a high potential for abuse and potentially severe psychological and/or physical dependence. As the drug schedule changes (Schedule II, Schedule III, etc.), so does the abuse potential—Schedule V drugs represents the least potential for abuse.

Scheduled drugs are some of the most powerful tools used in the treatment of chronic pain, anxiety, depression, and attention deficit disorders. For example, opiates are commonly prescribed for the management of acute or chronic pain despite research that long term biological efficacy is questionable [2]. To assist in the management of chronic pain, clinicians have opted for testing patients for compliance with their drug regimen. Routine assessment for non-compliance or non-medical use is frequently accomplished through urine drug testing (UDT) based on risk of drug misuse, abuse, and diversion. Although UDT is considered the common practice for detecting scheduled drug compliance, often patients are unable to provide a urine sample for various reasons. In this case, oral fluid drug testing (ODT) can serve as an effective alternative to UDT for medication monitoring [3].

Oral fluid drug testing is increasingly emerging as an alternative biological matrix for detecting drugs and monitoring patient medication compliance [4, 5, 6, 7, 8]. Moreover, in certain clinical situations clinicians may find oral fluid more beneficial for detection of specific drugs over UDT [9]. The matrix allows for easy collection, but attention to recovery, stability, and dilutions issues of some collection devices should be given consideration for pharmacokinetic studies [9, 10]. Although ODT opioid assays that use dilute-and-shoot methods with little sample manipulation have been developed and validated on AB SCIEX 4500 instruments with excellent calibration ranges (2.5-1,000 ng/mL) [11] robust ODT assays that also quantify the dextro (D-) and levo (L-) isomers of amphetamine and methamphetamine are less common. Furthermore, large ODT assays (50 drugs) have been successfully developed and validated on ultra-high performance triple quadrapole mass spectrometer (LC-MS/MS) [12], yet large ODT assay development on older LC-MS/MS technology and separation specificity are less common. This work highlights the development and validation of a fast, accurate, and inexpensive method to quantify 64 medications, illicit substances, and the D- and L- isomers of amphetamine and methamphetamine in human oral fluid specimens using liquid-liquid extraction and LC-MS/MS with an older model AB SCIEX 4000 instrument. Each sample was initially analyzed for 63 targeted analytes using LC-MS/MS and the same extracts injected a second time using a delta-9 tetrahydrocannabinol (THC) specific electrospray (ES) negative assay to detect the THC metabolite 11-Nor-9-carboxy-Δ9-tetrahydrocannabinol (THC-COOH). Then, for samples that initially showed positive for amphetamine or methamphetamine an additional sample was taken from the already extracted specimen and analyzed with a newly developed assay designed specifically to determine the D- and L- isomer status to define non-illicit versus illicit etiology. Accordingly, this work presents the development and validation of a robust human oral fluid drug assay—referred to in this paper as P63 assay. The assay development and validation are offered here for the benefit of high-throughput laboratories that seek novel solutions for a scoping ODT menu with fast and accurate chemical analysis using less expensive older model AB SCIEX instruments.

2. Results

2.1. Inter-day Average Back Calculated Calibration Standards

Each validation run contained calibration standards with theoretical concentrations of 1, 2.5, 5, 10, 25, 50, 100, 500, 1000 and 2000 ng/mL of each of the analytes with an additional negative run at 0.5 ng/mL. These line up with the P63 standard concentrations for ease of addition. The calibration curves were determined by plotting the theoretical concentration versus the area ratio for each standard. A weighted (1/x; where x=concentration) quadratic regression line was fit to the data and used to determine the concentration of unknown samples. Supplemental Table 1 shows the range of standard curves of the analytes and the correlation information. D- and L- curve concentrations were half the above concentration ranging from 0.25 (neg) to 1000 ng/mL. Mean R values were all at least 0.99 indicating good fit to the data.

2.2. Accuracy and Precision, LLOQ

Six replicates of each validation level were run over at least 3 days. The theoretical concentrations were 1, 5 or 25 for LLOQ on P63, 3, 15 or 75 for QC low, 600 ng/mL for QC mid and 800 or 1800 for the QC high values. The THC-COOH assay had concentrations of 1 ng/mL for the LLOQ, 3 ng/mL for the QC low, 600 for the QC mid and 2000 for the QC high. The D- and L- assay had an LLOQ of 2.5 ng/mL with a QC low of 7.5 ng/mL, a QC mid of 300 ng/mL and a high QC of 1000 ng/mL. Supplementary Tables 2 and 3 indicate inter-assay precision and accuracy were all below 20% except for 20.27 at the LLOQ of 1 for fentanyl. All other parameters for fentanyl were under 10% well under our 20% cutoff. The ranges for intra-assay variability and error are shown on Supplementary Table 4 and again fentanyl is the only one with percentages above 20. If only 2 of the 3 days are considered, the percentages drop below 20%, and are even below 15%.

2.3. Partial Volumes Accuracy and Precision

An MPA surrogate sample was prepared at 4000 ng/mL. To determine the concentration of this sample, a dilution must be made so the final concentration would be less than 2000 ng/mL to get it in the measurement range of the assay. Three replicates of four dilutions were made and tested: 1) 1:5 target 800 ng/mL; 2) 1:10 with a target of 400 ng/mL; 3) 1:20 with a target of 200 ng/mL; and 4) 1:50 with a target of 80 ng/mL. The results shown in Supplementary Table 5 indicate that a 1:10 dilution is safe for all analytes. Most of the analytes can be diluted at all levels. Notable exceptions are 7-aminoclonazepam and temazepam which only accommodated the 1:10 dilution.

2.4. Analyte Stability and Other Characteristics

QC samples were subjected to several conditions to test the stability of the analytes as shown in Supplementary Table 6.

2.5. Room Temperature, Refrigerator, and Freezer Stability

Samples with concentrations of 75 and 1800 ng/mL were prepared in triplicate. One set was kept at room temperature overnight (RT), a second set was kept in the refrigerator overnight (RF) and a third set was kept in the freezer overnight (FZ). These validation samples were then run and compared to a triplicate preparation of QC samples that had been analyzed as normal. All results show less than 20% deviation from expected.

2.6. Freeze/Thaw (FT) Stability

Validation samples with concentrations of 75 and 1800 ng/mL were frozen at -20 °C and thawed in sequence with samples taken after each FT cycle for a maximum of three cycles. These validation samples were analyzed in triplicate and compared to a triplicate preparation of validation samples that had not been subjected to this freeze/thaw cycle. The experimental results showed all meeting acceptance criteria.

2.7. Extracted Sample Stability

A stability experiment was performed where samples were stored in the instrument (3 day) or refrigerator (7 day) and re-injected after 3 and 7 days. All samples were within 20% of the initial results except for dihydrocodeine which was back within 20% on day 7.

2.8. Stability in Matrix

A series of triplicate samples were analyzed over 7 days for stability at room temperature, 4 °C and -20 °C. The results indicated that all analytes were stable for at least 7 days refrigerated and frozen. Most of the analytes were stable at room temperature except zopiclone (decreased) and ritalinic acid (increased) on day 7. A study of 30 days confirmed these characteristics for zopiclone and ritalinic acid. It also showed that chlordiazepoxide decreased after 14 days at room temperature as did methylphenidate. The decreased methylphenidate appeared to correspond with the increase in ritalinic acid concentration (metabolite of methylphenidate).

2.9. Matrix Effects and Recovery

Table 7 shows the effect of 10 different matrix lots tested by using a series of 75 ng/mL samples prepared in water, MPA and 10 different matrices. The results were acceptable with less than 20% CV across oral fluid, water and MPA meeting acceptance criteria. This is likely due to dilution in 1.5 mL Quantisal extraction buffer before extraction.

2.10. Selectivity

Multiple drugs that might have a potential for interfering with the assay analytes were run in the assay. Samples of 500 µL of 75 ng/mL QC were placed in a series of tubes to be run in triplicate. To the first set 50 µL of MeOH was added to act as the control. To the remaining tubes 50 µL of sample containing dextromethorphan, diphenhydramine, phenylephrine, salicylic acid, or combo (includes acetaminophen, caffeine, chlorpheniramine, ibuprofen, naproxen, and pseudoephedrine). These solutions were obtained from Cerilliant and were at a concentration of 1 mg/mL each except for the over the counter mix which was 100 µg/mL. Each solution was diluted to 20 µg/mL in methanol and this solution was used to spike samples as indicated above. Table 8 shows the results from this study. All samples met the acceptance criteria.

3. Discussion

The determination of prescription medications and illicit substances in oral fluids is one of the most non-invasive and easily observed sample collection methods. It provides a relatively simple and reliable means of sample collection coupled with a reduced chance of sample adulteration. Oral fluid also provides a viable alternative for measurement in patients that cannot provide an adequate urine sample volume such as catheterized patients. Analytes in oral fluid do not require a deconjugation step as in urine samples because the drugs are not glucuronidated for excretion in oral fluids and provides a lower cutoff than urine analysis. The only drawbacks of the oral fluid assay are that it has a shorter detection window and requires a more sensitive assay.

This paper presents a cost-effective means of analysis using older, less sensitive instruments (API SCIEX 4000) by using a liquid-liquid extraction method, concentration of the samples with a nitrogen dry-down, and a resuspension step. We validated the methods in accordance with FDA guidelines [13] with an LOQ of 5 ng/mL for most of the analytes except for fentanyl and THC-COOH at 1 ng/mL and gabapentin and pregabalin at 25 ng/mL. The ULOQ was 2000 ng/mL with a few exceptions that required a lower ULOQ of 1000 ng/mL. These were amitriptyline, amphetamine, chlordiazepoxide, ketamine, MDA, meperidine, methylphenidate, ritalinic acid, and tramadol.

4. Materials and Methods

4.1. Reagents and Standards

All analyte stock solutions at 1 mg/mL concentrations and deuterated internal standards at 100 µg/mL were purchased from Cerilliant Corporation (Round Rock, TX, USA). All organic solvents including methanol, acetonitrile, formic acid (88%), dichloromethane, 2 propanol and ethyl acetate were obtained from Fisher Scientific (Pittsburgh, PA, USA). Oral fluid Quantisal® extraction buffer and collection devices were obtained from Immunalysis Corporation (Pomona, CA, USA).

4.2. Mobile Phase and Extraction Solutions

Mobile phase A (MPA) solution was created with water, methanol, and formic acid (97.4:2.5:0.1) by using a 1L bottle to combine 974 mL of LC/MS grade water and 25 mL of LC/MS grade methanol; then 1 mL of formic acid (88%) was added and mixed thoroughly. The solution was stored at room temperature for up to two weeks. Mobile phase B (MPB) solution was created with acetonitrile and methanol (1:1) by using a 1L bottle to combine 500 mL of LC/MS-grade methanol and 500 mL of LC/MS-grade acetonitrile. This solution can be kept at room temperature for up to 1 year. A D- and L- mobile phase (MPDL) solution was created by adding ~993.2 mL of methanol to a 1L bottle. Then, using a mechanical pipette, 5 mL of type I water, 1.5 mL of acetic acid, and 0.3 mL of ammonium hydroxide were added and mixed thoroughly. This solution can be kept at room temperature for up to 1 year. A needle wash solution was created with methanol, acetonitrile, and clinical grade water (1:1:1) using a 1L bottle by adding equal volumes of methanol, acetonitrile, and water. Extraction solution 1 (ES1) was created with 50% dichloromethane and 50% 2-propanol by using a graduated cylinder under a fume hood. Equal volumes of dichloromethane and 2-propanol were added to a clean reagent bottle which was capped and mixed well. Extraction solution 2 (ES2) was created with 50% dichloromethane and 50% ethyl acetate by using a graduated cylinder under a fume hood. Equal volumes of dichloromethane and ethyl acetate were added to a clean reagent bottle which was capped and mixed well.

4.3. Standard Preparation

An 8000 ng/mL stock solution was made by combining analyte stock controls and diluting it with MPA. In contrast, D- and L- amphetamine and methamphetamine were added in an amount to make a 4000 ng/mL stock of each isomer so that combined they would produce an 8000 ng/mL solution of total amphetamine and methamphetamine. In contrast, this means that the range of the d and l SC is from 2.5 to 1000 ng/mL (half the concentration). The resulting stock standard was diluted with MPA to produce the standard curve (SC). Concentrations were 8000 (undiluted), 4000, 2000, 1000 400, 200, 100, 40, 20, 10, 4 and 2 ng/mL. These solutions were stored at the concentrations above. They underwent a dilution during the assay (1 part standard to 3 parts mobile phase and THC standard) to achieve the concentration desired in sample analysis with oral fluid (saliva). The standards and quality control (QC) were diluted (0.5 mL) with 1.5 mL of extraction buffer. This approximates the condition seen with saliva after collection with the Quantisal® oral fluid sample collection device. The final concentration in the 0.5 mL sample SC included the following points: 2000, 1000, 500, 250, 100, 50, 25, 10, 5, 2.5, 1 and 0.5 ng/mL.

The assay QCs were made similarly; first making a 7200 ng/mL spiking solution in MPA then diluting to 3200, 2400, 300, 60, 12, and 2 ng/mL. The D- and L- amphetamine and methamphetamine QCs were made at half concentrations. Final concentrations of each QC were 1800, 800, 600, 75, 15, 3 and 0.5 ng/mL, after the 1:4 dilution with MPA and THC QC same as the SC points noted above.

The internal standard working solution (ISWS) for the P63 assay was made by filling a 100 mL graduated cylinder to the 50 mL mark with 10% methanol in water and adding 250 µL of each of the internal standards listed above. The volume was brought to 100 mL with an additional 10% methanol producing a concentration of 250 ng/mL.

The 20000 ng/mL THC-COOH analyte stock solution was made by adding 200 µL of 1.0 mg/mL THC-COOH stock to a 15 mL polypropylene tube and bringing the volume to 10.0 mL using 50 % methanol in water. This was diluted further with 50% methanol in water to produce a SC with concentrations of 20000, 10000, 5000, 2500, 1000, 500, 250, 100, 50, 25, 10 and 5 ng/mL. Each solution underwent a dilution during the assay that resulted in final concentrations of 2000, 1000, 500, 250, 100, 50, 25, 10, 5, 2.5, 1 and 0.5 ng/mL.

The THC-COOH 18000 ng/mL quality control stock solution was made by adding 180 µL of 1.0 mg/mL THC-COOH stock to a 15 mL polypropylene tube and bringing the volume to 10.0 mL using 50 % methanol in water. This was diluted further with 50% methanol in water to produce a SC with concentrations of 18000, 8000, 6000, 750, 150, 30 and 5 ng/mL. These solutions underwent a dilution during the assay that results in final concentrations of 1800, 800, 600, 75, 15, 3 and 0.5 ng/mL.

The THC-COOH ISWS was made by diluting 300 µL of 100 µg/mL THC-COOH-D9 (Cerilliant) internal standard with 40 mL of 0.1 M ammonium acetate buffer pH 5 made by weighing out 7.7g of LC-MS/MS grade ammonium acetate and transferring to a 1-liter bottle, add ~700 mL of medical grade water that was capped and mixed until dissolved. The pH was adjusted to 5.0 with glacial acetic acid (3 shots of 733 µl) QS to 1L with medical grade water then capped and mixed well and stored at 2-8 C for up to 2 months.

4.4. Instrumentation

The liquid chromatography components of the LC-MS/MS system consisted of a model CBM-20A controller, 2 model Prominence LC-20AD pumps, a model DGU-20A5 degasser and a model SIL-20AC autosampler all obtained from (Shimadzu, Columbia MD, USA, based in Kyoto, Japan). The mass spectrometer used was a SCIEX API 4000 and the acquisition software was Analyst, v 1.5.2, build 5704 (Framingham, MA, USA). Nitrogen was obtained using a Peak ABN2ZA gas generator (Peak Scientific, Billerica, MA, USA). Reagents were weighed on a Mettler Toledo MX5 analytical micro balance (Fisher Scientific, Pittsburgh, PA, USA). Samples were dried on a TurboVap® LV (Uppsala, Sweden). Samples were vortexed on a Fisherbrand 120 multitube vortex. The analytical column was a Restek Ultra Biphenyl 5.0 µm (2.1 x 50 mm column), Catalog # 9109552 and the guard column was a Restek Ultra Biphenyl 5.0 µm (10 x 4 mm column) Catalog# 910950210 (Restek, Bellefonte, PA, USA) and Astec CHIROBIOTIC® V2 5.0 µm (2.1mm x 25 cm column) Catalog # 15020AST SUPLECO®, (Bellefonte, PA, USA).

4.5. Analyte Optimization

Individual analytes and internal standards were optimized by using T-infusion with 50% MPB mobile phase and tuning for declustering potential (DP), entrance potential (EP), collision energy (CE) and exit potential (CXP) at a flow rate of 0.7 mL/min. The two most abundant fragments were selected for monitoring using MRM. This resulted in the settings presented in Appendix A.

4.6. Sample Preparation and Procedures

Standards and QC were prepared in 13 x 100 mm tubes by adding 125 µl of P63 standard or QC, 50 µl of THC-COOH standard or QC, and 325 µl of MPA. A volume of 1.5 mL of Quantisal® buffer was added to all SC and QC preparations to match the volume of sample that is routinely extracted. A 2 mL aliquot of patient sample collected with the Quantisal® device was extracted along with the standards and QC in a vented biological safety cabinet and transferred to a properly labeled 13x100 mm glass tube. The ISWS for the P63 and THC-COOH assays were added to all tubes except for double blank and wash tubes. The samples were extracted with 2 mL of 50% dichloromethane: 50% 2 propanol. The samples were mixed with a mass vortex for 5 min. The tubes were transferred to centrifuge buckets, covered, and centrifuged 10 min at 3000 rpm (1690 x g) to facilitate separation. The clear bottom layer was transferred to a fresh 13 x 100 mm borosilicate glass tube. The remaining blue aqueous phase was extracted further with 2 mL of 50% dichloromethane and 50% ethyl acetate and mixed with a mass vortex for 5 min. The samples were centrifuged an additional 10 min at 3000 rpm (1690 x g) to fully separate the resultant layers. The bottom clear layer was combined with the corresponding clear aliquot from the first extraction step. The clear organic samples were dried under nitrogen in a TurboVap® LV.

Sample preparation for the LC-MS/MS was done by adding 125 µL of 50% MeOH and 50% water to each sample, SC and QC tube, wait for 10-15 min to dissolve, then vortex. Then 100 µL of the re-suspended sample was transferred to a corresponding well on the preparation plate. Next 400 µL of MPA was added to each well, the plate was covered with a plate mat and mixed briefly. The plate was centrifuged for 15 min at 4000 rpm (2272 x g) and samples were transferred (~300 µL) to a fresh, labeled plate. It was covered with a plate mat and analyzed for P63 and then THC-COOH.

Sample preparation for D- and L- analysis by LC-MS/MS involved transferring 50 µL of the already extracted standards, QC, and any samples of interest to a new plate. Then 450 µL of MPDL was added to each well and mixed with a multichannel pipette, the plate was covered with a plate mat and analyzed for the D- and L- isomers of amphetamine and methamphetamine.

The LC-MS/MS conditions and separation parameters for all three methods/assays used in this study are expanded in Tables 9-11. In summary samples were injected with the P63 parameters followed by a second injection with the THC-COOH parameters. A 50 µL aliquot of these prepared samples, standards and quality controls were transferred to a fresh plate and diluted with 450 µL of D- and L- assay mobile phase for LC-MS/MS injection to determine D- and L- isomer concentrations.

4.7. Method Validation Procedures
4.7.1. Matrix lot-to-lot comparison

Individual lots of human matrix (saliva) differ according to a person’s overall health and hydration status [14]. A single lot of oral fluid is not enough to demonstrate the ruggedness of the assay system when such variability in the matrix exists [14]. Due to this, and in accordance with current CAP standards, a minimum of 10 lots of human matrix were collected from donors who verify that they are not taking the analytes that are being validated. These donor oral fluid samples were spiked at a low-level concentration with each analyte. These samples were prepared, extracted, and analyzed as described above. The responses were calculated and the analyte to internal standard (IS) ratio and %CV is shown in Table 7.

4.7.2. Analytical measurement range

The analytical measurement range (AMR) of the assay refers to the concentration range that the assay is validated within and is determined by running a series of calibration curve standards covering a concentration range that encompass the concentration of analyte expected to find in patient samples [15]. The limits of the AMR were bounded by the lower limit of quantitation (LLOQ) and the upper limit of quantitation (ULOQ). The dynamic range may be described by a linear or quadratic fit [16, 17]. Calibration curves were created using a minimum of six non-zero calibration points. To be accepted as the AMR, all points describing the calibration curve must pass within ± 20% of the nominal concentration [13]. Furthermore, the correlation coefficient (R2) for the calibration curve must be ≥ 0.98, or R should be ≥0.99 to be acceptable [18, 19].

4.7.3. Sensitivity

The sensitivity of the assay system refers to the ability to reliably produce a signal throughout the entire calibration range, but specifically at the low-end of the calibration curve [20]. In hyphenated mass spectrometry assays, a signal that produces a signal to noise ratio (S/N) of ≥10 is considered valid for the LLOQ of an assay system [21]. Further, a S/N ratio of ≥5 is considered clear enough for the limit of detection. We tested the sensitivity of the assay system by injecting 6 replicates of the LLOQ over 3 days and evaluating the resulting analytical determinations. Standard acceptance criteria of ±20% of nominal concentration apply.

4.7.4. Intra-day precision and accuracy

Intra-day precision and accuracy were determined using six replicates of each of three quality control (QC) sample determinations and LLOQ from across at least three validation runs. Concentrations of the QC samples ranged across the curve, with the low QC set at approximately 3 times the LLOQ or less, the mid QC near the middle of the linear range and the high QC set at 80-90% of the ULOQ. Percent accuracy was determined for each individual measurement using the equation:

| VdVn Vn |×100%

Where Vd is the concentration determined from the calibration curve and Vn is the nominal concentration for the QC standard. Precision was determined for each standard level by first determining the standard deviation of the six replicate standards and then applying the following equation:

SD Vm *100%

Where SD is the standard deviation of the six replicates and Vm is the mean value of the standard. To be accepted, the precision and accuracy for the replicate determinations must be ≤20% at each level.

4.7.5. Inter-day precision and accuracy

Inter-day precision and accuracy were determined using all replicates of each of three quality control (QC low, QC mid, and QC high) and LLOQ sample determinations from the analytical runs performed on 3 separate days. Concentrations of the QC samples ranged across the curve, with the low QC set around 3 times the LLOQ, the mid QC near the middle of the linear range, and the high QC set at 80-90% of the ULOQ.

4.7.6. Exogenous interfering substances

Drugs that are known or suspected of interfering with similar bioanalytical systems should be evaluated to ensure that they do not suppress ionization or cause false-positive results for a given analyte [22, 23]. The following medications were evaluated: over-the-counter mix (consisting of acetaminophen, ibuprofen, pseudoephedrine, caffeine, and naproxen), salicylic acid, phenylephrine, phentermine, diphenhydramine, and dextromethorphan. A high concentration of the possible interfering drug (typically 2,000 ng/mL or greater) was spiked into a low QC sample (15 – 75 ng/mL low QC). Acceptance criteria for a substance to be deemed as non-interfering is that the quantitated value for the low QC should be within ± 20% of the nominal value [24]. Furthermore, the spiked substance should not cause a false-positive or a false-negative result.

4.7.7. Exogenous interfering substances

A spiked solution was created at a concentration above the ULOQ in this case 4000 ng/mL. The sample was run at discrete dilutions 1:5, 1:10, 1:20, and 1:50. Concentration determinations for all dilutions should be within ± 20% of the nominal value following correction for the dilution factor [24, 25]. More recent literature suggests that the signal to noise ratio of both the quantification trace and the qualifying ion trace be 3-10 [26, 27]. On occasion, an analyte will not have a qualifying ion that passes this criterion while still permitting the quantification trace to remain in a meaningful range. These instances should be documented in the laboratory SOP or validation report.

4.7.8. Carryover

Carryover is the presence of an analyte in a blank injection following a positive injection, resulting in a false-positive sample [28]. The injection needle should be washed in-between samples with a needle wash solution that is intended to remove contamination from the surface of the needle. The efficiency of this process is monitored during validation by assessing carryover in the following manner. Samples are injected in the following sequence: high QC, wash, high QC, wash, high QC, wash. Peak areas are integrated for both the analyte and internal standard. Peak area in the wash solutions should be 0.1 % or less of that found in the High QC standard. In addition, the mean of the peak area in the three wash solutions following the high QC replicates should be less than 20% of the LLOQ being used for the assay [29].

5 Conclusion

The novelty of this study is a sensitive and low-cost method of ODT developed and validated for the determination of 64 drug analytes and the D- and L- isomers of amphetamine and methamphetamine from the same liquid-liquid extraction on an older model API 4000 mass spectrometer. Separation and detection are based on 3 LC-MS/MS injections. This method has an 8 minute run-time for most of the analytes, 1.5 minutes for the THC-COOH, and 10 minutes for the D- and L- isomeric separation for amphetamine and methamphetamine. The assay is quite sensitive for the majority of the analytes with a cutoff of 5 ng/mL except for fentanyl and THC-COOH at 1 ng/mL and gabapentin and pregabalin at 25 ng/mL, The assay also has good precision and accuracy and would add a valuable option to high throughput laboratories seeking a robust testing alternative to UDT methods and for medical providers seeking to achieve medication compliance [30, 31].

Supplementary Materials: The following supporting information can be downloaded at: Table S1-S7.

Author Contributions: For research articles with several authors, a short paragraph specifying their individual contributions must be provided. The following statements should be used “Conceptualization, B.R. and R.C.; methodology, B.R. and J.P.; validation, B.R., J.P., and M.L.; formal analysis, M.L.; data curation, J.P. and B.R.; writing—original draft preparation, R.C. and B.R.; writing—review and editing, R.C. All authors have read and agreed to the published version of the manuscript.

Funding: This research received no external funding.

Data Availability Statement: https://doi.org/10.7910/DVN/CQTWE1, Harvard Dataverse.

Conflicts of Interest: The authors have no conflicts of interest to report concerning this study; however, the authors are either employed by, or work on behalf of Advanta Genetics, the laboratory that developed and validated this assay.

References

  1. United States Drug Enforcement Administration. Drug Scheduling. https://www.dea.gov/drug-information/drug-scheduling. Accessed May 20, 2022.
  2. Van Ree JM, Gerrits MA, Vanderschuren LJ. Opioids, reward and addiction: an encounter of biology, psychology, and medicine. Pharm Rev 1999, 51, 341-396.
  3. Heltsley R, DePriest A, Black DL, et al. Oral fluid drug testing of chronic pain patients II. Comparison of paired oral fluid and urine specimens. J of Anal Tox. 2012;36(2):75-80. https://doi.org/10.1093/jat/bkr019[CrossRef] [PubMed]
  4. Wagner E, Raabe F, Martin G, et al. Concomitant drug abuse of opioid dependent patients in maintenance treatment detected with a multi‐target screening of oral fluid. The Am J on Addict. 2018;27(5):407-412. https://doi.org/10.1111/ajad.12737[CrossRef] [PubMed]
  5. Pil K, Verstraete A. Current developments in drug testing in oral fluid. Thera Drug Mon. 2008;30(2):196-202. https://doi:10.1097/FTD.0b013e318167d563[CrossRef] [PubMed]
  6. Desrosiers NA, Huestis MA. Oral fluid drug testing: analytical approaches, issues and interpretation of results. J of Anal Tox. 2019; 43(6):415-443. https://doi.org/10.1093/jat/bkz048[CrossRef] [PubMed]
  7. Robbins B, Carpenter RE, Long M, Perry J. A Human Oral Fluid Assay for D-and L-Isomer Detection of Amphetamine and Methamphetamine Using Liquid-Liquid Extraction. J of Anal Meth in Chem. 2022;4819599. https://doi.org/10.1155/2022/4819599[CrossRef] [PubMed]
  8. Robbins, B., Long, M., Perry, J., Carpenter, R. E. Analysis of D- and L- isomers of (meth)amphetamine in human K2EDTA plasma. J. of Biomed and Life Sci. 2023;(3)5343:):1-12. https://doi.org/10.31586/jbls.2023.534[CrossRef]
  9. Langel K, Engblom,C, Pehrsson A, et al. Drug testing in oral fluid—evaluation of sample collection devices. J of Anal Tox. 2008;32(6):393-401. https://doi.org/10.1093/jat/32.6.393[CrossRef] [PubMed]
  10. White RM, Moore CM. Oral fluid pharmacokinetics. In: Thomas B, ed. Detection of Drugs and Their Metabolites in Oral Fluid. 1st ed. Amsterdam: Elsevier; 2018:11–39[CrossRef]
  11. West R, Mikel C, Hofilena, D, Guevara M. Positivity rates of drugs in patients treated for opioid dependence with buprenorphine: a comparison of oral fluid and urine using paired collections and LC-MS/MS. Drug and Alcoh Dep. 2018;193:183-191. https://doi.org/10.1016/j.drugalcdep.2018.07.023[CrossRef] [PubMed]
  12. Arantes ACF, da Cunha KF, Cardoso MS, Oliveira KD, Costa JL. Development and validation of quantitative analytical method for 50 drugs of antidepressants, benzodiazepines and opioids in oral fluid samples by liquid chromatography–tandem mass spectrometry. Forensic Tox. 2021;39(1);179-197. https://doi.org/10.1007/s11419-020-00561-8[CrossRef]
  13. Food and Drug Administration. Guidance for Industry: Bioanalytical Method Validation, US Department of Health and Human Services, FDA, Center for Drug Evaluation and Research, Rockville, MD, 2018.
  14. Caporossi L, Santoro A, Papaleo B. Saliva as an analytical matrix: state of the art and application for biomonitoring. Biomarkers. 2010;15(6):475-487. https://doi.org/10.3109/1354750X.2010.481364[CrossRef] [PubMed]
  15. Gjerde H, Langel K, Favretto D, Verstraete, AG. Detection of illicit drugs in oral fluid from drivers as biomarker for drugs in blood. Forensic Sci International. 2015;256:42-45. https://doi.org/10.1016/j.forsciint.2015.06.027[CrossRef] [PubMed]
  16. Killeen AA, Long T, Souers R, Styer P, Ventura CB, Klee GG. Verifying performance characteristics of quantitative analytical systems: calibration verification, linearity, and analytical measurement range. Arch of Path & Lab Med. 2014;138(9):1173-1181. https://doi.org/10.5858/arpa.2013-0051-CP[CrossRef] [PubMed]
  17. Rule GS, Clark ZD, Yue B, Rockwood AL. Correction for isotopic interferences between analyte and internal standard in quantitative mass spectrometry by a nonlinear calibration function. Anal Chem. 2013;85(8):3879-3885. https://doi.org/10.1021/ac303096w[CrossRef] [PubMed]
  18. Asuero AG, Sayago A, González AG. The correlation coefficient: an overview. Critical Rev in Anal Chem. 2006;36(1):41-59. https://doi.org/10.1080/10408340500526766[CrossRef]
  19. Van Loco J, Elskens M, Croux C, Beernaert H. Linearity of calibration curves: use and misuse of the correlation coefficient. Accreditation and Qual Assurance. 2002;7(7):281-285. https://doi.org/10.1007/s00769-002-0487-6[CrossRef]
  20. Viswanathan CT, Bansal S, Booth B, et al. Quantitative bioanalytical methods validation and implementation: best practices for chromatographic and ligand binding assays. Pharma Research. 2007;24(10):1962-1973. https://doi.org/10.1007/s11095-007-9291-7[CrossRef] [PubMed]
  21. Meyer C, Seiler P, Bies C, Cianciulli C, Wätzig H, Meyer VR. Minimum required signal‐to‐noise ratio for optimal precision in HPLC and CE. Electrophoresis. 2012;33(11):1509-1516. https://doi.org/10.1002/elps.201100694[CrossRef] [PubMed]
  22. Langel K, Engblom,C, Pehrsson A, et al. Drug testing in oral fluid—evaluation of sample collection devices. J of Anal Tox. 2008;32(6):393-401. https://doi.org/10.1093/jat/32.6.393[CrossRef] [PubMed]
  23. Liang HR, Foltz RL, Meng M, Bennett P. Ionization enhancement in atmospheric pressure chemical ionization and suppression in electrospray ionization between target drugs and stable‐isotope‐labeled internal standards in quantitative liquid chromatography/tandem mass spectrometry. Rapid Comm in Mass Spect. 2003;17(24):2815-2821.[CrossRef] [PubMed]
  24. Gosetti F, Mazzucco E, Zampieri D, Gennaro MC. Signal suppression/enhancement in high-performance liquid chromatography tandem mass spectrometry. J of Chromatography A. 2010;1217(25):3929-3937. https://doi.org/10.1016/j.chroma.2009.11.060[CrossRef] [PubMed]
  25. Peters FT, Remane D. Aspects of matrix effects in applications of liquid chromatography–mass spectrometry to forensic and clinical toxicology—a review. Anal and Bioanal Chem. 2012;403(8):2155-2172. https://doi.org/10.1007/s00216-012-6035-2[CrossRef] [PubMed]
  26. Viswanathan CT, Bansal S, Booth B, et al. Quantitative bioanalytical methods validation and implementation: best practices for chromatographic and ligand binding assays. Pharma Research. 2007;24(10):1962-1973. https://doi.org/10.1007/s11095-007-9291-7[CrossRef] [PubMed]
  27. Mannu RS, Turpin PE, Goodwin L. Alternative strategies for mass spectrometer-based sample dilution of bioanalytical samples, with particular reference to DBS and plasma analysis. Bioanalysis. 2014;6(6);773-784.[CrossRef] [PubMed]
  28. Cai HL, Wang F, Peng WX et al. Quantitative analysis of erythromycylamine in human plasma by liquid chromatography-tandem mass spectrometry and its application in a bioequivalence study of dirithromycin enteric-coated tablets with a special focus on the fragmentation pattern and carryover effect. J of Chromatography B. 2014;947:156-163. https://doi.org/10.1016/j.jchromb.2013.12.019[CrossRef] [PubMed]
  29. Sheehan TLY, Yost RA. What’s the most meaningful standard for mass spectrometry: instrument detection limit or signal-to-noise ratio. Curr Trends Mass Spectrom. 2015;13:16-22.
  30. Carpenter RE, Silberman D, Takemoto JK. Transforming prescription opioid practices in primary care with change theory. Health Serv Insights. 2021;14. https://doi.org/10.1177/11786329211058283[CrossRef] [PubMed]
  31. Paul AK, Smith CM, Rahmatullah M, Nissapatorn V, Wilairatana P, Spetea M, Gueven N, Dietis N. Opioid analgesia and opioid-induced adverse effects: A review. Pharmaceuticals. 2021;14(11):1091. https://doi.org/10.3390/ph14111091[CrossRef] [PubMed]

Copyright

© 2024 by authors and Scientific Publications. This is an open access article and the related PDF distributed under the Creative Commons Attribution License which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Article Metrics

Citations
Google Scholar 2

If you find this article cited by other articles, please click the button to add a citation.

Article Access Statistics
Article Download Statistics
Article metrics
Views
259
Downloads
57
Citations
2

How to Cite

Robbins, B., Carpenter, R. E., Long, M., & Perry, J. (2023). Quantifying 64 drugs, illicit substances, and D- and L- isomers in human oral fluid with liquid-liquid extraction. Open Journal of Medical Sciences, 3(1), 1–38. Retrieved from https://www.scipublications.com/journal/index.php/ojms/article/view/591
  1. United States Drug Enforcement Administration. Drug Scheduling. https://www.dea.gov/drug-information/drug-scheduling. Accessed May 20, 2022.
  2. Van Ree JM, Gerrits MA, Vanderschuren LJ. Opioids, reward and addiction: an encounter of biology, psychology, and medicine. Pharm Rev 1999, 51, 341-396.
  3. Heltsley R, DePriest A, Black DL, et al. Oral fluid drug testing of chronic pain patients II. Comparison of paired oral fluid and urine specimens. J of Anal Tox. 2012;36(2):75-80. https://doi.org/10.1093/jat/bkr019[CrossRef] [PubMed]
  4. Wagner E, Raabe F, Martin G, et al. Concomitant drug abuse of opioid dependent patients in maintenance treatment detected with a multi‐target screening of oral fluid. The Am J on Addict. 2018;27(5):407-412. https://doi.org/10.1111/ajad.12737[CrossRef] [PubMed]
  5. Pil K, Verstraete A. Current developments in drug testing in oral fluid. Thera Drug Mon. 2008;30(2):196-202. https://doi:10.1097/FTD.0b013e318167d563[CrossRef] [PubMed]
  6. Desrosiers NA, Huestis MA. Oral fluid drug testing: analytical approaches, issues and interpretation of results. J of Anal Tox. 2019; 43(6):415-443. https://doi.org/10.1093/jat/bkz048[CrossRef] [PubMed]
  7. Robbins B, Carpenter RE, Long M, Perry J. A Human Oral Fluid Assay for D-and L-Isomer Detection of Amphetamine and Methamphetamine Using Liquid-Liquid Extraction. J of Anal Meth in Chem. 2022;4819599. https://doi.org/10.1155/2022/4819599[CrossRef] [PubMed]
  8. Robbins, B., Long, M., Perry, J., Carpenter, R. E. Analysis of D- and L- isomers of (meth)amphetamine in human K2EDTA plasma. J. of Biomed and Life Sci. 2023;(3)5343:):1-12. https://doi.org/10.31586/jbls.2023.534[CrossRef]
  9. Langel K, Engblom,C, Pehrsson A, et al. Drug testing in oral fluid—evaluation of sample collection devices. J of Anal Tox. 2008;32(6):393-401. https://doi.org/10.1093/jat/32.6.393[CrossRef] [PubMed]
  10. White RM, Moore CM. Oral fluid pharmacokinetics. In: Thomas B, ed. Detection of Drugs and Their Metabolites in Oral Fluid. 1st ed. Amsterdam: Elsevier; 2018:11–39[CrossRef]
  11. West R, Mikel C, Hofilena, D, Guevara M. Positivity rates of drugs in patients treated for opioid dependence with buprenorphine: a comparison of oral fluid and urine using paired collections and LC-MS/MS. Drug and Alcoh Dep. 2018;193:183-191. https://doi.org/10.1016/j.drugalcdep.2018.07.023[CrossRef] [PubMed]
  12. Arantes ACF, da Cunha KF, Cardoso MS, Oliveira KD, Costa JL. Development and validation of quantitative analytical method for 50 drugs of antidepressants, benzodiazepines and opioids in oral fluid samples by liquid chromatography–tandem mass spectrometry. Forensic Tox. 2021;39(1);179-197. https://doi.org/10.1007/s11419-020-00561-8[CrossRef]
  13. Food and Drug Administration. Guidance for Industry: Bioanalytical Method Validation, US Department of Health and Human Services, FDA, Center for Drug Evaluation and Research, Rockville, MD, 2018.
  14. Caporossi L, Santoro A, Papaleo B. Saliva as an analytical matrix: state of the art and application for biomonitoring. Biomarkers. 2010;15(6):475-487. https://doi.org/10.3109/1354750X.2010.481364[CrossRef] [PubMed]
  15. Gjerde H, Langel K, Favretto D, Verstraete, AG. Detection of illicit drugs in oral fluid from drivers as biomarker for drugs in blood. Forensic Sci International. 2015;256:42-45. https://doi.org/10.1016/j.forsciint.2015.06.027[CrossRef] [PubMed]
  16. Killeen AA, Long T, Souers R, Styer P, Ventura CB, Klee GG. Verifying performance characteristics of quantitative analytical systems: calibration verification, linearity, and analytical measurement range. Arch of Path & Lab Med. 2014;138(9):1173-1181. https://doi.org/10.5858/arpa.2013-0051-CP[CrossRef] [PubMed]
  17. Rule GS, Clark ZD, Yue B, Rockwood AL. Correction for isotopic interferences between analyte and internal standard in quantitative mass spectrometry by a nonlinear calibration function. Anal Chem. 2013;85(8):3879-3885. https://doi.org/10.1021/ac303096w[CrossRef] [PubMed]
  18. Asuero AG, Sayago A, González AG. The correlation coefficient: an overview. Critical Rev in Anal Chem. 2006;36(1):41-59. https://doi.org/10.1080/10408340500526766[CrossRef]
  19. Van Loco J, Elskens M, Croux C, Beernaert H. Linearity of calibration curves: use and misuse of the correlation coefficient. Accreditation and Qual Assurance. 2002;7(7):281-285. https://doi.org/10.1007/s00769-002-0487-6[CrossRef]
  20. Viswanathan CT, Bansal S, Booth B, et al. Quantitative bioanalytical methods validation and implementation: best practices for chromatographic and ligand binding assays. Pharma Research. 2007;24(10):1962-1973. https://doi.org/10.1007/s11095-007-9291-7[CrossRef] [PubMed]
  21. Meyer C, Seiler P, Bies C, Cianciulli C, Wätzig H, Meyer VR. Minimum required signal‐to‐noise ratio for optimal precision in HPLC and CE. Electrophoresis. 2012;33(11):1509-1516. https://doi.org/10.1002/elps.201100694[CrossRef] [PubMed]
  22. Langel K, Engblom,C, Pehrsson A, et al. Drug testing in oral fluid—evaluation of sample collection devices. J of Anal Tox. 2008;32(6):393-401. https://doi.org/10.1093/jat/32.6.393[CrossRef] [PubMed]
  23. Liang HR, Foltz RL, Meng M, Bennett P. Ionization enhancement in atmospheric pressure chemical ionization and suppression in electrospray ionization between target drugs and stable‐isotope‐labeled internal standards in quantitative liquid chromatography/tandem mass spectrometry. Rapid Comm in Mass Spect. 2003;17(24):2815-2821.[CrossRef] [PubMed]
  24. Gosetti F, Mazzucco E, Zampieri D, Gennaro MC. Signal suppression/enhancement in high-performance liquid chromatography tandem mass spectrometry. J of Chromatography A. 2010;1217(25):3929-3937. https://doi.org/10.1016/j.chroma.2009.11.060[CrossRef] [PubMed]
  25. Peters FT, Remane D. Aspects of matrix effects in applications of liquid chromatography–mass spectrometry to forensic and clinical toxicology—a review. Anal and Bioanal Chem. 2012;403(8):2155-2172. https://doi.org/10.1007/s00216-012-6035-2[CrossRef] [PubMed]
  26. Viswanathan CT, Bansal S, Booth B, et al. Quantitative bioanalytical methods validation and implementation: best practices for chromatographic and ligand binding assays. Pharma Research. 2007;24(10):1962-1973. https://doi.org/10.1007/s11095-007-9291-7[CrossRef] [PubMed]
  27. Mannu RS, Turpin PE, Goodwin L. Alternative strategies for mass spectrometer-based sample dilution of bioanalytical samples, with particular reference to DBS and plasma analysis. Bioanalysis. 2014;6(6);773-784.[CrossRef] [PubMed]
  28. Cai HL, Wang F, Peng WX et al. Quantitative analysis of erythromycylamine in human plasma by liquid chromatography-tandem mass spectrometry and its application in a bioequivalence study of dirithromycin enteric-coated tablets with a special focus on the fragmentation pattern and carryover effect. J of Chromatography B. 2014;947:156-163. https://doi.org/10.1016/j.jchromb.2013.12.019[CrossRef] [PubMed]
  29. Sheehan TLY, Yost RA. What’s the most meaningful standard for mass spectrometry: instrument detection limit or signal-to-noise ratio. Curr Trends Mass Spectrom. 2015;13:16-22.
  30. Carpenter RE, Silberman D, Takemoto JK. Transforming prescription opioid practices in primary care with change theory. Health Serv Insights. 2021;14. https://doi.org/10.1177/11786329211058283[CrossRef] [PubMed]
  31. Paul AK, Smith CM, Rahmatullah M, Nissapatorn V, Wilairatana P, Spetea M, Gueven N, Dietis N. Opioid analgesia and opioid-induced adverse effects: A review. Pharmaceuticals. 2021;14(11):1091. https://doi.org/10.3390/ph14111091[CrossRef] [PubMed]

Citations of